Specification

PDF Download
Code Range Accuracy Resolution Data transmission
9501-1200 0~1250μm ±(3%L+1)μm (range≤1250μm), ±(3%L+10)μm (range>1250μm) 0.1μm (range<100μm), 1μm (range≥100μm) Type-C
9501-1200N 0~1250μm ±(3%L+1)μm (range≤1250μm), ±(3%L+10)μm (range>1250μm) 0.1μm (range<100μm), 1μm (range≥100μm) Type-C

View the complete
product specifications

FAQ

How can I choose between FE and NFE probes? Can they be used interchangeably?

The FE probe or NFE probe cannot be used interchangeably. Please choose according to the substrate and coating.

● Magnetic induction probe (FE):

Substrate: iron, steel, magnetic stainless steel

Coating: zinc, copper, chrome, tin, plastic, powder, paint (Not suitable for non-magnetic stainless steel substrates or nickel coatings)

● Eddy current probe (NFE):

Substrate: copper, aluminum, zinc, non-magnetic stainless steel

Coating: plastic, powder, paint, anodizing (Not suitable for chrome or zinc plating)

Can a flat calibration foil be used for curved workpieces?

No. It is recommended to calibrate the instrument using a standard film sample on an uncoated substrate sample with the same curvature before taking measurements.

Can I measure a high-temperature workpiece directly?

Not recommended. High temperature may affect the probe and cause measurement errors. Please allow the workpiece to cool to room temperature before measurement.

Will measuring too close to the edge affect the reading?

Yes. If the probe is too close to the edge, the magnetic field or eddy current may leak at the edge and cause inaccurate readings. Keep the measuring point far from the edge.